Quality Assurance Guidelines for
Testing Using the OraQuick® Rapid HIV-1 Antibody
Test
Use of trade names and commercial
sources is for identification only and does not imply endorsement by the
Centers for Disease Control and Prevention or the U.S. Department of
Health and Human Services.
The
following CDC staff members prepared this report:
Devery
Howerton, Ph.D. Thomas L. Hearn, Ph.D. Division of Laboratory
Systems Public Health Practice Program Office
Ida M.
Onorato, M.D. Bernard M. Branson, M.D. Naomi Bock,
M.D. Division of HIV/AIDS Prevention National Center for HIV, STD
and TB Prevention
Jonathan E. Kaplan, M.D. Mark A. Rayfield,
Ph.D. Division of AIDS, STD, and TB Laboratory Research National
Center for Infectious Diseases
in consultation
with
Elliot Cowan,
Ph.D. Steven I. Gutman, M.D. Food and Drug
Administration Rockville, MD
Nelson Michael, M.D.,
Ph.D. United States Army Walter Reed Army Institute of
Research Rockville, MD
Chris Alderidge National
Alliance of State and Territorial AIDS Directors Washington,
D.C.
Patricia Charache, M.D. Johns Hopkins Medical
Institutions Baltimore, MD
Jane Getchell, Dr.
PH Delaware Public Health Laboratory Smyrna,
DE
Gregg Gonsalves Gay Men's Health Crises New
York, NY
Patrick Keenan, M.D. University of
Minnesota Minneapolis, MN |
Carlyn Collins,
M.D. Eileen Burke Public Health Practice Program Office and
National Center for HIV, STD and TB Prevention, CDC
Judy
Yost Valerie Coppola Centers for Medicare & Medicaid
Services Baltimore, MD
Catherine Johnson,
M.T. COLA Columbia, MD
Liisa Randall,
Ph.D. Michigan Dept. of Community Health Okemos,
MI
Karen Kroc Cook County Bureau of Health
Services Chicago, IL
Belinda Yen-Lieberman,
PhD Cleveland Clinic Cleveland, Ohio
Marlene
LaLota, MPH Florida Department of Health Tallahassee,
FL |
Introduction and Background
Purpose This document provides guidance on
quality assurance (QA) practices for sites using or planning to use the
OraQuick® Rapid HIV-1 Antibody Test to detect antibodies to the
human immunodeficiency virus (HIV).
Background The OraQuick Rapid HIV-1 Antibody
Test is the first rapid HIV point-of-care (i.e., testing and results are
available in one visit) test approved by the U.S. Food and Drug
Administration (FDA). It is also the first test for HIV that the FDA has
waived under the Clinical Laboratory Improvement Amendment regulations
(CLIA). The OraQuick test uses whole blood obtained from puncture of a
finger. Results are available within 20 to 60 minutes.1
Positive results with the OraQuick rapid test are preliminary, however,
and must be followed up with an acceptable confirmatory test. Although the
OraQuick test device is simple to use and can provide reliable results
when the manufacturer’s directions are followed, mistakes can occur at any
point in the testing process. To reduce mistakes and to ensure that the
FDA restrictions for sale of the test are followed (see Appendix A for
information on the FDA sales restrictions), a site must have a QA program
in place before offering OraQuick testing. The guidelines in this document
outline the basic parts of a QA program.
How these guidelines were developed These
guidelines were developed after many discussions on quality assurance for
rapid HIV testing within the Centers for Disease Control and Prevention
(CDC) and culminated from the discussions at a meeting of experts convened
by the CDC at the end of January 2003. This working group included
individuals from Federal agencies—CDC, FDA, U.S. Department of Defense
(DOD), and the Centers for Medicare & Medicaid Services (CMS)—as well
as individuals outside the Federal government with expertise in rapid
point-of-care testing, QA, HIV prevention programs, and private and public
health laboratories.
How to use these guidelines This document
outlines the basic processes and procedures that should be in place before
a site offers rapid HIV testing. It describes steps that can be taken to
identify and prevent errors in the testing process. Because the OraQuick
test will be used in many different settings, each site needs to decide
how to fit the various QA elements into its own workflow and system of
operation. For example, following these guidelines in a large clinic or
hospital environment where on-site laboratory support is available may be
quite different from using them in a small voluntary counseling and
testing site or outreach setting with few staff and resources. These
guidelines are intended to assist a range of providers in developing
policies, processes and procedures to ensure high quality HIV testing
services.
How this document is organized This document
includes text and appendices that provide basic information that staff in
sites offering OraQuick testing should know. It includes information
on:
- The basics of a QA program for testing using the
OraQuick test.
- An overview of government rules that apply to using
this test.
- Examples of forms/checklists that can be used to keep
track of QA outcomes.
Basic Elements of a Quality Assurance Program
What is quality assurance? Quality assurance
(QA) refers to planned, step-by-step activities that let one know that
testing is being carried out correctly, results are accurate, and mistakes
are found and corrected to avoid adverse outcomes. Quality assurance is an
ongoing set of activities that help to ensure that the test results
provided are as accurate and reliable as possible for all persons being
tested. Quality assurance activities should be in place during the entire
testing process; this means from the time a person asks to be tested using
the rapid HIV test to providing the test result.
How does quality assurance differ from quality
control? As described above, QA is an overall program of activities
throughout the entire testing process. Quality control (QC) is one part of
the QA program. See Quality
Control for details on quality control testing for the OraQuick test.
Here are definitions for both terms2:
|
Term |
Definition and
activities performed |
| Quality
assurance |
Planned
and organized activities to help ensure that certain requirements
for quality will be met |
| Quality
control |
Operational techniques or tasks that are in place to find and
correct problems that might occur
Basic |
Basic elements of a QA program for OraQuick Rapid HIV-1
Antibody testing Even though the OraQuick test is simple to use,
things can go wrong. To help find and prevent problems, the basic elements
of a QA program should be in place before offering testing. These basic
elements are the building blocks of a QA program and are listed below.
More detail on these five elements is provided in this
document.
- Organization of the QA program
- Testing personnel
- Process control
- Before testing
- During testing
- After testing
- Documents and records
- Troubleshooting
Organization of the QA
program
Establishing a QA program Resources are needed
to establish and maintain a QA program, no matter how simple. Someone must
oversee the program and ensure the necessary staff and supplies are
available. Each organization must:
- Identify the person(s) responsible for managing the QA
program (this could be a senior staff member, outside consultant or a
network of individuals who oversee different aspects of the QA
program).
- Write procedures (step-by-step instructions) and make
them available to all staff involved in testing (see the list of
recommended procedures below).
- Verify the testing process (see below).
- Ensure staff know how to perform processes and
procedures (see the section on personnel
who conduct testing).
- Create mechanisms for communication so that those who
need to know are informed about QA issues, as well as all staff, when
appropriate.
- Develop and implement mechanisms to ensure the site
meets all applicable Federal, State, and other regulatory requirements.
Each site offering testing must have a CLIA Certificate of Waiver if
they are performing only the OraQuick test or the OraQuick test and
other waived tests, or be included under an organization with a CLIA
exception for limited public health or mobile testing. Each site must
also meet Federal requirements for biohazard safety, as well as
applicable State rules. See Appendix A for more information on
regulatory requirements.
Verifying the testing process Before offering
the test to clients or patients, each site should make sure (verify) that
the testing process works as planned. This verification should be
completed before testing is offered. Verification includes ensuring that
staff have been trained and are able (competent) to perform their assigned
tasks, the test kits work as expected (e.g., make sure the test gives
accurate results for a referenced panel of non-reactive, weakly reactive
and reactive specimens), and the logistics for providing confirmatory
testing (if a person tests positive, he or she still has to have a test to
confirm the finding) and biohazardous waste handling are in
place.
Providing written procedures It is strongly
recommended that step-by-step, written instructions be made available to
all staff performing testing. This will help to ensure that personnel know
how to perform specific tasks and testing success is not left to chance.
Testing personnel must follow instructions provided by the manufacturer.
Additional procedures, as listed below, should be provided along with the
manufacturer’s instructions. Text from the current OraQuick package insert
may be used for some of the items denoted by an asterisk (*) in the list
below. Written instructions should describe how to:
- Train new employees, assess their ability to do the
testing and document training.
- Provide information to persons being tested before
testing.*
- Use gloves and other personal protective
equipment.
- Safely dispose of biohazardous waste, including used
lancets.
- Maintain sufficient supplies and unexpired test and
control kits, follow the manufacturer’s instructions for storage, and
check performance of new test kit lots and shipments with external
controls (see section on "Checking
inventory and test kits").
- Maintain and document the temperature of the room and
refrigerator where the tests and controls are stored and testing is
performed.
- Perform quality control testing and take action (e.g.,
contact the manufacturer) if controls don’t work.
- Collect the OraQuick specimen. *
- Perform steps in the test procedure. *
- Report results.
- Refer specimens or persons being tested for
confirmatory testing and manage confirmatory test results.
- Record test and quality control results.
- Conduct external quality assessment (see "External
assessment").
- Review records and store and destroy them when they are
outdated (how long test result records are kept as part of a medical
record may be subject to State or other requirements).
- Troubleshoot and take corrective action when things go
wrong.
Testing
Personnel
Overview Having qualified, trained staff to
perform and supervise OraQuick testing and the various activities in the
QA program is one of the most important factors for ensuring accurate and
reliable results. Key aspects of this element include:
- Qualifications
- Training
- Competency assessment (i.e., how well they are doing
their job)
Personnel qualifications Since the OraQuick test
is waived under CLIA, there are no specific Federal requirements on who
can perform the test. Each site should find out if there are State or
other requirements for personnel that they must meet. Beyond any
regulatory requirements, it is recommended that certain qualities be
considered when selecting personnel to perform the OraQuick test. The
following list of qualities resulted from practical considerations and
expert opinion:
- Sincerity and commitment – A dedication to
performing testing according to defined procedures.
- Literacy – The ability to read instructions and
record results is critical.
- Organizational skills – The need for this
quality will depend on the number and complexity of tasks an individual
performs in the testing process. If test volume is high and the
individual performing testing is doing several tests or managing several
other tasks simultaneously, organizational skills can be
critical.
- Decision-making skills – Testing personnel
should be able to interpret results and be able to recognize and handle
problems that might come up.
- Communication skills – If the person performing
the test also is the one who shares results or other information with
the person being tested, being able to communicate clearly is
important.
Components of training Training is crucial to
ensuring quality testing.3 Training is also required to be able
to purchase the OraQuick test kit (see Appendix A for details on the FDA
sales restrictions). Staff should be fully trained on how to perform their
assigned tasks and responsibilities. Training should be documented for
each staff member; using training checklists is one way to handle this
documentation (see Appendix B for an example of a training checklist). The
key components to include in a training program are:
- How to perform the test, including procedures performed
before, during and after testing.
- How testing is integrated into the overall counseling
and testing program.
- The importance of QA and the elements of the site’s QA
program.
- The use and importance of Universal (or Standard)
Precautions/biohazard safety.
Training method Experience with training to perform the OraQuick test (CDC
unpublished data) shows that a training method should optimally include
the following activities:
- Read the instructions for performing the test.
- Watch someone perform the test or view a video of
someone performing the test.
- Practice performing the test with positive and negative
control materials.
- Practice performing the finger-stick collection
procedure.
- Review the procedures and forms on how to document
testing.
Competency assessment Before a trainee is
permitted to perform testing alone for the first time, his or her ability
to conduct the test should be demonstrated and documented. This
assessment should also be carried out at periodic intervals after
training, such as every six months or other interval as determined by the
testing site. This assessment can be carried out in many ways, but
regardless of the method, every task for which a staff member is
responsible should be evaluated. A supervisor or trainer should
perform the assessment, using a combination of methods to determine
competency. Examples of these methods are presented below.
Assessing performance of tasks done before
testing To assess the task performance before testing, staff should
be observed as they:
- Check and record the temperatures of the testing and
storage areas.
- Set up the testing area, label the device and prepare
control and test results log sheets.
- Run the external controls and record results.
Assessing performance of tasks done during
testing To assess staff’s ability to perform the test and interpret
results
- Observe the staff member performing the finger-stick,
collecting the blood on a test loop and placing it into the testing
vial.
- Observe how the test is performed on a client/patient.
If such observation will interfere with actual client-provider
interactions, observe test performance on a volunteer.
- Evaluate the use of Universal or Standard Precautions
and procedures for biohazard and sharps (e.g., lancets, needles) waste
disposal.
- Review results obtained on a panel of referenced
specimens that show a range of results, such as five specimens that
include non-reactive, weakly reactive and reactive results. Control
materials supplied by the manufacturer may be used as a source of
specimens in the panel. In addition, specimens may be obtained from
laboratories performing confirmatory testing or from other commercial
sources.
- Appraise the individual’s ability to interpret results.
This might include using previously used test devices or pictures of
devices that show non-reactive, weakly reactive, reactive and invalid
results.
Assessing performance of tasks done after
testing To assess task performance after testing:
- Review test records and quality control results
documentation.
- Observe oral reporting of results to a test subject (if
trainee’s responsibility).
- Observe venous blood and/or oral fluid specimen
collection and handling for confirmatory testing. If the frequency of
OraQuick reactive results is low, the trainee should be observed
collecting blood and/or oral fluid from a staff volunteer and
demonstrate how it is processed for confirmatory testing.
- Verify that confidentiality is maintained.
Process
Control
What is process control? Process control refers
to the activities and techniques that are carried out to ensure that the
testing procedures are performed correctly, the environment is suitable,
and the test kit works as expected to produce accurate and reliable
results.
Steps in the testing process Steps in the
testing process follow the path of workflow beginning with tasks before
testing, followed by those conducted during and after testing. This path
of workflow and the associated steps are shown in the table below.
Detailed descriptions about each of the steps listed in this table are
provided in the remainder of this document.

|
Before testing
|
During testing
|
After
testing |
-
Check storage and room
temperatures daily
-
Check inventory and
test kit lots, as needed
-
Receive request for
testing
-
Provide HIV/AIDS
information to the test subject
-
Set up test area, label
test device
-
Perform external
quality control according to the manufacturer’s and the site’s
instructions |
|
-
Clean up and dispose of
biohazardous waste
-
Report results to
client
-
Document
results
-
Collect, process and
transport confirmatory test specimens
-
Manage confirmatory
test results
-
Participate in external
quality assessment (periodically)
|
Before
Testing
Overview As shown in the table
above, there are a number of steps that must be followed before testing
the blood sample for HIV. These activities are in place to ensure that the
conditions in which the tests are stored and performed are suitable, the
test area and the test subject are prepared, and the test is working
appropriately.
Temperature control: test kits and control
kits Test kits and controls must be stored in an environment within
the temperature ranges specified by the manufacturer. Store test kits at
2° to 27° C (35° to 80° F). If test kits are refrigerated, the pouch
containing the test device and developer solution must be brought to room
temperature (15° to 27° C or 59° to 80° F) before opening. Control kits
must be refrigerated at 2° to 8° C (35° to 46° F). To ensure these
temperature ranges are maintained, monitor and document temperatures of
the storage areas each day testing is performed. If the temperature falls
outside of the specified range, take action as needed to adjust the
temperature. To monitor the temperatures, place thermometers in the
storage areas (e.g., in the refrigerator and on the shelf in the room
where kits are stored). Check and record temperatures on a log sheet each
day testing is performed. An example temperature log is provided in
Appendix C.
Temperature control: testing area The
temperature in the area where the test will be performed must be within
the range of 15° to 27° C (59° to 80° F). If the test must be performed at
a temperature below 15°C/59°F or above 27°C/80°F, run external controls
that have been stored within the proper temperature range to find out if
the test can be performed at another temperature (see the section below on
external controls). If testing is carried out in the field, monitor the
temperature of the test and control kits in their portable storage
containers and check the temperature where testing will be performed if it
appears to be outside the specified range. If there are doubts about the
testing area temperature or whether test kits have stayed within the
appropriate temperature range, run a positive and negative external
control as described in the quality control section below.
Checking inventory and test
kits Procedures should be in place to ensure that an adequate
supply of unexpired test kits, controls, and supplies is available. Test
kits and controls have a defined shelf life. Use the oldest first. Never
use test or control kits beyond their expiration dates. It is helpful to
use a log sheet to document when test and control kits are received, their
lot numbers and expiration dates. Also, once control vials are opened,
they are stable for 21 days. Therefore, record on the vial the date it is
opened and discard unused opened controls after 21 days. As described in
the package insert and in the section on quality control below, run the
positive and negative controls with new lots and new shipments of test
kits before using them for testing, to verify that they work as
expected.
Setting up the testing area and labeling the test
device Before testing, the testing area should be prepared
according to the specific site procedure, which should include directions
for setting up the workspace listed in the test kit instructions, as well
as instructions for how to label testing devices and complete report
forms, including the method for identifying each person to be tested to
ensure specimens are not mixed up during the testing process. Labeling is
especially important when more than one test is being performed at the
same time. Label components of the test with the name or identifying
number of the person being tested before collecting the specimen. These
components include the developer solution vial, test device, and documents
for recording results. Using preprinted labels improves the efficiency of
performing this task.
Note: Do not place a label over the two holes on
the back of the test device as this can cause an invalid
result.
Providing information to test subjects OraSure Technologies,
Inc., provides a “Subject Information” pamphlet that must be given to each
person getting tested prior to performing the HIV rapid test. Each site
may provide additional information. For further details, see the CDC
website http://www.cdc.gov/hiv/pubs/rt-counseling.htm,
the Revised Guidelines for HIV Counseling, Testing, and Referral,
MMWR Recommendations and Reports, RR-19, vol. 50, November 9,
20014 and applicable State or local rules.
Quality control There are two
types of quality control (QC) for the OraQuick test. These are described
in the table below.
|
Type of quality
control |
Description of
activity |
|
Internal
Controls |
A control is built in
to each testing device to verify that the specimen was adequate and
the solution flowed through the device as
intended. |
|
External
Controls |
Known reactive and
non-reactive specimens (controls) are available from the
manufacturer to sites purchasing the OraSure Rapid HIV Test. They
are used to evaluate the accuracy of the test in detecting antibody
to HIV and to check if the person conducting the test performs it
correctly. |
External quality control To verify that the test
device is accurately detecting HIV-1 antibodies, external positive and
negative controls must be tested from time to time. The test kit
manufacturer provides external controls in the form of the OraQuick Rapid
HIV-1 Antibody Test Kit Controls. This control kit must be ordered
separately from the test kit. It includes one vial each of an HIV
antibody-negative (non-reactive) and positive (reactive) human plasma
control. How often controls are run to verify the accuracy of the test
will depend on the number of tests carried out by the site, how often new
test kit shipments or lot numbers are received by a site, changes in how
the tests are stored and testing area temperatures, and how often staff
who conduct the testing change. An example of a log for control testing
results is available in Appendix D.
Run external controls according to the manufacturer’s
instructions The manufacturer has set guidelines for the minimum
number of times to run the negative and positive controls. This is
described in the test kit instructions, which specifies running controls
under the following circumstances:
- By each new operator prior to performing testing on patients,
- When opening a new test kit lot (a test kit lot is defined as the
boxes of test devices that contain either 25 or 100 tests that have the
same lot number labeled on the outside of the boxes),
- Whenever a new shipment of test kits is received (even if it is the
same kit lot number in current use),
- If the temperature of the test storage area falls outside of 2°-27°
C (35°-80° F),
- If the temperature of the testing area falls outside of 15°-27° C
(59°-80° F), and
- At periodic intervals as dictated by the user facility.
Frequency of running external controls on the basis of
test volume In addition to the specific circumstances listed in the
manufacturer’s instructions, testing sites should determine the optimal
frequency for running controls on the basis of their test volume. When
external controls provide incorrect results, none of the tests that were
run since the last time control results were correct can be considered
valid. This means that everyone who was tested since the last time
controls ran correctly will need to be called back and retested (unless a
confirmatory test was ordered). Sites testing large numbers of persons,
and especially those that offer anonymous testing, should plan to run
controls more often than facilities that conduct fewer tests. Each site
needs to decide how often to run controls based on its own situation and
testing practices. Instructions for some other waived tests recommend
running external controls each time a new box of 25 tests is opened.
Facilities that test 25 or more subjects a day should run controls every
day. Low volume sites, such as those testing fewer than 25 subjects per
month, should run external controls every two to four weeks at a minimum.
Controls should be run more often if new lots or shipments are opened or
if storage or testing temperatures fluctuate.
During Testing
Overview This phase of the testing process
involves running the test and interpreting the results. Activities during
testing include collecting the specimen, performing the test, interpreting
the internal control and client/patient test results, and following
biohazard safety guidelines.
Collecting the OraQuick specimen Follow the written procedure
for finger-stick specimen collection. Further information on collecting
blood by skin puncture can be found in Procedures and Devices for the
Collection of Diagnostic Blood Specimens by Skin
Puncture.5
Performing the test and interpreting
results Follow the manufacturer’s instructions for performing the
test and interpreting the results. Results can be one of the
following:
- Nonreactive (negative)
- Reactive (preliminary positive)
- Invalid (the test result is inconclusive and cannot be
interpreted; see below for information on handling invalid results)
Evaluating internal control results Each
OraQuick device includes a built-in (internal) control. When an
appropriate line develops at the center of the “C” location on the device,
the patient’s specimen has been correctly loaded and traveled through the
test strip, indicating a valid test. Additional information is provided in
the test kit package insert. These controls are included in every device,
and control results are evaluated with every test. If the internal control
does not produce the expected result, the test result for the patient is
not valid, cannot be reported, and the test must be repeated. If a
second invalid result occurs, external controls should be evaluated as
described below before repeating the test a third time.
Running external controls to troubleshoot invalid
results CDC experience (unpublished data) has shown that external
controls should be run to help find out if repeated invalid test results
are due to the test device, test performance, or the patient specimen. If
the same test kit lot yields repeated invalid results, the test kit may
have gone bad. It is important to run the positive and negative controls
whenever two consecutive invalid test results are obtained on a person
being tested.
Biohazard safety/Universal (Standard) Precautions All
specimens and materials contacting specimens must be handled as if they
are capable of transmitting an infectious organism. As described in
Appendix A, each site must ensure that the Occupational Safety and Health
Administration (OSHA) bloodborne pathogens standards are met, that is
persons doing the testing must know how to safely handle potentially
infectious specimens. Also, according to Universal (Standard) Precautions,
all human blood should be treated as if known to be infectious for HIV,
hepatitis B virus, and other bloodborne pathogens. Sites must have
available and follow procedures for biohazard safety to include
instructions for the use of gloves, hand washing, sharps and biohazardous
waste disposal, spill containment and disinfection. A different pair of
gloves should be worn for collecting a specimen from each person being
tested. Used gloves should be handled as biohazardous waste. For further
details on these precautions see the OraQuick package insert, OSHA
regulations and guidelines on Universal and Standard Precautions. 1,
6, 7, 8
After Testing
Overview Quality assurance extends to those
activities completed following the performance of the test. Each site
should have established procedures for:
- Reporting and recording results,
- Referring specimens (or test subjects, if specimens are
not collected on-site) for confirmatory testing,
- Managing confirmatory test results, and
- Conducting external quality assessment.
Reporting results Reporting procedures should
describe how results are provided to the person being tested (verbal
and/or written results) and how results are documented in the person’s
chart and in the test result logs. Some States have laws and regulations
that include certain reporting criteria for HIV testing results. Check
with your State agency for more information on these requirements. See
Appendix A for State agency contact information and Appendix E for an
example of a test result log.
Referral for confirmatory testing Whenever the
OraQuick test result is reactive (preliminary positive), a confirmatory
test must be performed to confirm that the person being tested is infected
with HIV. Therefore, each site must have established procedures for
referral of either test specimens or persons being tested for confirmatory
testing when OraQuick results are reactive. If specimens are collected
on-site, the site must establish procedures describing how to collect,
label, process, store and document specimen transfer; transport the
confirmatory test specimens to the site(s) where they will be tested; and
obtain the confirmatory results to give to the client/patients. It should
be indicated on the specimen transfer sheet that the specimen is from an
individual who had a reactive OraQuick rapid test result. See the Appendix
F for an example of a specimen transfer sheet. Collecting confirmatory
specimens on-site may improve follow-up, since some clients may not go
elsewhere for the testing or to obtain results. However, if the site is
not able to collect confirmatory test specimens, a procedure must be in
place for referring persons to another site to obtain this
testing.
Confirmatory testing protocols For confirmatory
testing, the current standard testing algorithm should be followed, with
the following exceptions:
- All OraQuick reactive (preliminary positive) results
must be followed up with either a Western blot or immunofluorescent
assay (IFA) for confirmation.
- Confirmatory testing can be done on blood (plasma,
serum or dried blood spots) or oral fluid specimens. Urine testing
should not be performed due to its lower sensitivity (i.e., ability to
detect positive results).
- With blood specimens, enzyme immunoassay (EIA)
screening tests prior to the Western blot or IFA confirmatory test are
optional. If an EIA is performed, even if it is non-reactive, the
specimen must proceed to Western blot or IFA testing (reactive EIA
specimens will automatically be tested by Western blot or IFA). For oral
fluid testing, both EIA and Western blot testing should be performed to
confirm results.
Follow up testing for negative confirmatory
result Most confirmatory test results will be positive; however,
some may be negative or indeterminate. If the confirmatory test result is
negative, specimen mix-up needs to be ruled out versus a false positive
OraQuick result. If the Western blot or IFA test is negative, it is
recommended that:
- For blood specimens, a confirmatory test should be
repeated with a new specimen to rule out specimen mix-up.
- For oral fluid specimens, a repeat confirmatory test
with a blood specimen should be done, since the oral fluid test is less
sensitive than the blood test.
Follow up testing for indeterminate
results Occasionally, confirmatory test results are indeterminate.
If the Western blot or IFA is indeterminate, it is recommended
that:
- For blood specimens, the person should be advised to
return for repeat testing in one month. See CDC’s Revised Guidelines for
HIV Counseling, Testing and Referral found at http://www.cdc.gov/mmwr/preview/mmwrhtml/rr5019a1.htm.
- For oral fluid specimens, the Western blot or IFA test
should be repeated using a blood specimen.
Managing confirmatory results OraQuick testing
sites that refer specimens for confirmatory testing should have
established procedures describing how to:
- Match the client’s/patient’s confirmatory test results
with their OraQuick results to find potential discrepancies and to
ensure that testing was performed according to the protocol described
above,
- Report the test result to the person being tested,
and
- Obtain any additional specimens needed to resolve
potential specimen mix-up and for retesting, as needed.
Handling result discrepancies Procedures should
describe how to handle result discrepancies when the OraQuick result was
reactive and the confirmatory test negative or indeterminate. If the
laboratory providing confirmatory testing performed an EIA test only and
reported a non-reactive or negative result, the OraQuick testing site
should contact the confirmatory testing laboratory and request a Western
blot test or IFA test. If the original specimen is not available, a new
specimen will need to be collected from the person in question to be used
for confirmatory testing.
External
assessment External assessment, or an evaluation of the testing
process by a source outside the testing site, can look at how testing is
being performed and whether it is being performed reliably. It can also
help to identify existing or potential problems. Moreover, information
gathered can provide an educational tool to improve performance. Some form
of external assessment is highly recommended, but it is not required by
Federal (CLIA) regulations since the test is waived and the test kit
manufacturer does not specifically require it.
Methods for external assessment Every reactive
OraQuick test is externally assessed by a second, confirmatory test.
However, if there is a low prevalence of HIV infection in the population
being testied, these assessments may be rare and will not provide an
external check for the majority of the results, i.e., those that are
nonreactive. Other ways to assess performance may be needed. Some external
assessment mechanisms include:
- Comparing the OraQuick reactive results with the
confirmatory test results.
- Arranging for someone outside the organization to
observe testing.
- Participating in a proficiency testing or external
evaluation program (for more information on these programs, see Appendix
G).
Documents and
Records
Overview One of the hallmarks of a QA program is
comprehensive documentation. Sites using the OraQuick test should have
policies and procedures describing what QA records are required and how
and when they are reviewed, stored and destroyed. Having a supervisor
review records periodically is recommended. State regulations or other
governmental or accrediting agencies may require facilities to have
specific record retention policies. QA records include the
following:
- Training documentation (Appendix B)
- Temperature logs (Appendix C)
- External control result logs (Appendix D)
- Test result logs (Appendix E)
- Specimen transfer logs (Appendix F)
Temperature logs Temperature logs should include
a daily record of the refrigerator temperature in which controls are
stored, the temperature where test kits are stored and the temperature of
the testing area. Thermometers should be placed in each location.
Laboratory grade thermometers (can be purchased from medical or laboratory
supply houses) are recommended and their accuracy checked periodically
(e.g., every six months) by comparison with another
thermometer.
External control result logs External control
records should include the date and time of control testing, lot number
and expiration of the test kit, lot number and expiration date of the
controls, control results, and corrective action taken if control results
are unacceptable. Control records should be kept in the order in which
they were completed so they can be easily compared with the test records.
This will help find answers if there are questions about testing performed
within a specific time frame.
Test result logs Test result records should
include the date and time of testing, an identifier for the person being
tested, a test kit lot number and expiration date, test result, action
taken if the result was invalid, identification of the person who
performed the test, whether confirmatory testing was requested, including
the type of specimen sent for confirmation (e.g., oral fluid, blood), and
the confirmatory test results when they are available. If more than one
person is conducting testing, there should be a mechanism to
chronologically link the test record log sheets to detect problems, such
as invalid results occurring repeatedly with the same kit lot
number.
Troubleshooting
Overview Each site should have a method to
detect and resolve problems that occur at any point in the testing
process, especially those that may affect the accuracy of test results.
Significant problems should be immediately reported to the appropriate
supervisory personnel.
Procedures Procedures should be available to all
testing personnel for the following:
-
When to
discontinue testing, e.g., when the external control results are
unacceptable as described in the package insert.
-
How to take
corrective action, or an action taken in response to a problem, such as
contacting the manufacturer when the external control results are
unacceptable and following the advice provided.
-
How to
document problems and actions taken, such as a logbook where problems
and corrective actions taken can be recorded.
-
How to
verify the corrective actions taken addressed the problem.
References
- OraQuick Rapid HIV-1 Antibody Test package insert.
OraSure Technologies, Inc., Bethlehem, PA 18015, 2003.
- NCCLS document HS1-A. A quality system model for health
care, Approved guideline. NCCLS, Wayne, PA, 2002.
- NCCLS document GP21-A.Training verification for
laboratory personnel, Approved guideline. NCCLS, Wayne, PA, 1995.
- CDC. CDC revised guidelines for HIV counseling,
testing, and referral. MMWR Recommendations and Reports.
2001;RR-19:50.
- NCCLS document H4-A4. Procedures and devices for the
collection of diagnostic blood specimens by skin puncture, Approved
guideline. NCCLS, Wayne, PA, 1999.
- Occupational Safety and Health Administration
regulations, 29CFR Part 1910. Available from http://www.osha.gov/SLTC/bloodbornepathogens/index.html
- CDC. Perspectives in disease prevention and health
promotion update: Universal precautions for prevention of transmission
of human immunodeficiency virus, hepatitis B virus, and other bloodborne
pathogens in health-care settings. MMWR 1988; 37(24):377-88.
- Garner JS, Hospital Infection Control Practices
Advisory Committee. Guideline for isolation precautions in hospitals.
Infect Control Hosp Epidemiology 1996;17:53-80, and Am J Infect Control
1996;24:24-52.
Appendixes:
Government Regulations and Sample Forms (PDF only - 407 KB, 9
pages)
LEGEND: = Link is outside of the DHAP domain...click the BACK
button to return to this page.
Adobe Acrobat (TM) Reader v5.0
or higher needs to be installed on your computer in order to
read the QA Guidelines for OraQuick in PDF. Download the Adobe Acrobat (TM) Reader
Added on: August 8, 2003 Centers
for Disease Control & Prevention National Center for HIV, STD, and
TB Prevention Divisions of
HIV/AIDS Prevention Please send comments/suggestions/requests to:
dhaprti@cdc.gov
|